FOR SECURITY PURPOSES - because Internet Explorer is no longer supported by Microsoft, we suggest that you interact with our secure site through one of our supported browsers - Google Chrome, Firefox, or MS Edge. If you continue to use this website with Internet Explorer you do so at your own risk and you may encounter problems.
Blog Hero Background

How to Reduce Your HPLC Analysis Time

Blog
Fast HPLC Methods: How to Reduce Your HPLC Analysis Time
Shorten HPLC analysis time with smarter column choices, optimized gradients, and efficient method design, faster analysis without losing resolution.

Posted on Aug 10, 2022

Reducing HPLC analysis time isn’t just about running shorter methods; it’s about smarter method design. Optimized methods help laboratories process more samples, reduce solvent use, and boost productivity without compromising separation quality. With the right HPLC column, efficient method design, and a strong understanding of system variables, you can significantly decrease analysis time and keep your workflows running at peak performance.

Why Shorter HPLC Analysis Times Matter

Today’s analytical labs operate under tight deadlines and heavy workloads. Shortening analysis time increases sample throughput, frees valuable instrument capacity, and supports faster decision-making across R&D, QC, and manufacturing workflows.

Faster run times also reduce mobile-phase consumption and thus money savings. But speed must never undermine resolution. The goal is a balance delivering high-throughput chromatographic performance while maintaining sharp peaks and system suitability standards.

Factors That Influence HPLC Method Speed

Several method parameters directly control how quickly an HPLC separation runs. Optimizing these allows labs to design fast HPLC methods without compromising chromatographic result.

Reduce Column Length for Faster Analysis

Shorter columns naturally reduce run time by shortening the separation path. Moving from a 150 mm to a 100 mm or 50 mm format can accelerate your method dramatically. Combined with high-efficiency particle technology, shorter columns maintain outstanding resolution and throughput.

Use Smaller or Core-Shell Particles for Higher Efficiency

Smaller or core-shell particles provide higher separation efficiency by reducing the mass transfer and the Eddy diffusion contribution to the van Deemter equation. This allows higher flow rates or shorter columns while maintaining resolution.

Increase Mobile Phase Flow Rate Carefully

Flow rate strongly affects HPLC run time. Increasing flow accelerates elution and shortens analysis, but higher flow also raises backpressure, which must remain within system limits. According to the Van Deemter equation, low flow increases longitudinal diffusion, while high flow amplifies mass transfer resistance, both reducing efficiency. The optimal flow rate balances speed and resolution without exceeding pressure constraints.

More Ways to Achieve Shorter HPLC Run Times Without Sacrificing Quality

Optimize Gradient Programs for Faster Elution

Gradient methods significantly reduce analysis time compared to isocratic runs. By gradually increasing the mobile phase strength during the separation, strongly retained compounds elute faster without requiring an extended isocratic hold. This approach allows early-eluting compounds to separate under mild conditions while accelerating the elution of late peaks, shortening the overall run time and improving peak shape for complex mixtures.

Increase Column Temperature to Reduce Viscosity

Higher column temperatures lowers mobile-phase viscosity, reducing backpressure and allowing faster flow rates for shorter run times. It's important to keep in mind that temperature also affects retention and selectivity, as well as compound stability and column durability.

Reduce Extra-Column Volume

Extra-column volume refers to the dead volume outside the column, including tubing, fittings, and detector cell. While it does not directly change the intrinsic speed of elution, it affects peak dispersion. This impact is minimal for long columns but becomes critical for short columns and high-speed methods, where even small dead volumes can significantly degrade resolution. Minimizing extra-column volume is essential to preserve efficiency in ultra-fast separations.

Use UHPLC-Capable Systems When Possible

UHPLC systems tolerate higher pressures than conventional HPLC, allowing the use of smaller particles and higher flow rates without exceeding system limits. This capability enables faster separations with improved efficiency compared to standard HPLC, which is restricted by lower pressure thresholds.

Improve Sample Cleanup to Simplify Separations

Sample preparation does not alter the intrinsic chromatographic speed but can indirectly affect it. Poorly prepared samples may introduce particulates or incompatible solvents, increasing system backpressure or causing unstable baselines, which can force slower flow rates or longer equilibration times. Conversely, clean, well-prepared samples allow operation at optimal flow rates and gradient conditions, enabling faster and more reliable analysis.

Summary: Balancing Speed and Quality

Shortening HPLC analysis time requires the right combination of column technology, thoughtful method optimization, and awareness of system contributions. Faster run times boost productivity and lower operating costs, but only when balanced with resolution, robustness, and data integrity. Labs can achieve fast HPLC methods that keep pace with modern analytical demands by pairing advanced particle designs with optimized gradients, minimized dispersion, and strong sample preparation.

FAQs

How can I reduce HPLC run time quickly?

Shorten the column, use high-efficiency particles, and refine gradient slopes. Ensure all adjustments are supported by good HPLC testing practice.

Can I increase flow rate without affecting results?

Yes, you can increase the flow rate to speed up the analysis, provided the system can handle the higher backpressure and the column maintains acceptable efficiency. Always check pressure limits and monitor resolution to ensure performance is not compromised.

Does higher temperature speed up HPLC?

Yes. Increasing column temperature reduces viscosity, allowing faster flow rates and improved peak shapes. This can shorten run time without compromising separation quality.